Switch the solvents to your appropriate solvents
-
Take control of the Chromeleon Plus Software (if not already) and make sure the appropriate solvent lines (A, B, C or D) are selected in the Eleuent Box and change the lines you aren’t using to 0%
-
Loosen the purge value (this only takes about a turn and a half, just loosen it until you feel slight tension, DO NOT LOOSEN ALL THE WAY)
-
You will hear a clicking noise as it purges
-
Once it is done purging you will see this message in the event log: Pump has stopped purging. Please close the purge valve.