there's 4 settings for calibration
- left TC gain
- left TC offset
- right TC gain
- right TC offset
the formula in code is
left = (left * leftgain) + leftoffset
right = (right * rightgain) + rightoffset
act = (left + right) / 2
so in theory you should be able to calibrate the oven by running a profile, and dumping the values via serialport, while at the same time measuring with a known good temperature sensor taped to the pcb
here's some code that does that with a Uni-T ut181a
https://gist.github.com/aep/03db148e01796316018ca241f6825d88
remember to set gains to 1 and offsets to zero before doing that, so you can then match the curves in matlab/octave using the above formula
for some reason in reality the offset is different tho. i had to set them to half the value than in matlab to get a good curve. some experimentation is required