Expand Up @@ -36,19 +36,21 @@ Siemens scanners come with a useful program called ``ideacmdtool`` that can conf

Configure the default settings (these steps should be performed **BEFORE** registering a patient):

.. image:: network_online_export.jpg

4. Enter ``4`` to go to the "Online export defaults" menu
5. Enter ``1`` and set "Target port" to ``-1``
6. Enter ``3`` and set "Target path" to ``y:`` (or whatever path you mounted the SAMBA folder on)
7. Enter ``8`` to set ``SendBuffered`` to ``OFF``
8. Enter ``q`` to go back to the main ``ideacmdtool`` menu

.. image:: network_online_export.jpg

Configure the session settings (these steps should be performed **AFTER** registering a patient, since these flags get reset every time you register a new patient):

.. image:: network_switches.jpg

9. Enter ``5`` to go to the "Switches" menu
10. Enter ``8`` to set ``SendIMA`` to ``ON``
11. Enter ``q`` to go back to the main ``ideacmdtool`` menu
12. Enter ``q`` to exit ``ideacmdtool``

.. image:: network_switches.jpg