…ut_complete from sensitivity_value.
Want to check this one last time at the beamline, then it will be ready for review. |
Tests ran with errors here. |
You mean locally? The CI looks like it's just that weird decrease in test coverage. |
Meant to say: "Tests ran with no new errors here." |
Finally had a chance to test this at the beamline. Totally didn't forget about it and just noticed it was still in draft while I was looking for something else. Anyway, I think it works, best as I can tell. I swapped the gain back and forth a hundred or so times in a while loop, and they always got set in the right order. |
Any thoughts about merging this PR? |
There was a problem hiding this comment.
Choose a reason for hiding this comment
The reason will be displayed to describe this comment to others. Learn more.
LGTM
@canismarko - You can merge now |
* Added edge_align plan for beam edge alignment * Need to find a better way to check signal quality. * Formatted using ruff * small formatting changes * removed unused stats import * removed another unused import * Added a check function that yields correct results * intermediate commit saving progress * finished try/except when bad signal is detected * Fixed formatting * Added scipy to environment.yaml & requirements.txt * CI #909 switch to 'ruff' * DOC #909 update release notes * MNT #909 revise per 'ruff check' * CI #909 update * Bump davidslusser/actions_python_ruff from 1.0.0 to 1.0.1 Bumps [davidslusser/actions_python_ruff](https://github.com/davidslusser/actions_python_ruff) from 1.0.0 to 1.0.1. - [Release notes](https://github.com/davidslusser/actions_python_ruff/releases) - [Commits](davidslusser/actions_python_ruff@v1.0.0...v1.0.1) --- updated-dependencies: - dependency-name: davidslusser/actions_python_ruff dependency-type: direct:production update-type: version-update:semver-patch ... Signed-off-by: dependabot[bot] <support@github.com> * MNT #872 hoist from MPE devices * TST #872 adjust for new hoist * MNT #872 * MNT #872 refactor * MNT #872 hoist utils * DOC #912 * ENV #938 bump the minimum requirement * DOC #932 update relase notes * DOC #938 contribution from @sveseli * MNT #914 setup.py --> pyproject.toml * DOC #921 copyright year * MNT #914 * Added edge_align function for beam edge alignment * added scipy to pyproject.toml * CI #909 switch to 'ruff' * DOC badge * DOC badge * DOC #940 example scans with sscan record * DOC #940 it's only a word ... * DOC #940 * DOC #940 it's a demo, actually * DOC #940 * DOC #940 convert the blocking call to st.wait() * DOC #940 per review * DOC #940 per review * DOC #940 proofreading * DOC #940 per review, remove the section with the polling loop * DOC #940 new title * DOC #940 more proofing * DOC #940 clarify the default sscan record array size of 1,000 * DOC #940 refactor to polling loop. Show the recommended style in a details section. * DOC #940 * Added put_complete=True to sensitivity value and unit for the SRS-570 Pre-amp. * Added a test for the sensitivity_unit put_complete, and removed the put_complete from sensitivity_value. * CI #955 move requirements file * MNT #955 sphinx config * DOC #955 re-arrange * DOC #955 apply to lower-level pages * DOC #955 add docs upload * CI #955 * CI #955 * CI #955 wip * CI #955 * CI #955 * DOC #917 release note * DOC #957 merge overview into home page * DOC #955 * DOC #955 * DOC #955 * DOC #955 * Added edge_align function for beam edge alignment * Added edge_align function for beam edge alignment * Added edge_align function for beam edge alignment * DOC #909 update release notes * DOC #932 update relase notes * Added edge_align function for beam edge alignment * CI #909 switch to 'ruff' * DOC #955 * Fixed merge issue * fixed unused import * change for tests to pass * Formatting changes * Fixed repeated section * Small formatting changes * Small syntax change * Added toolz & scipy to conf.py --------- Signed-off-by: dependabot[bot] <support@github.com> Co-authored-by: Codrea <ecodrea@nefarian.xray.aps.anl.gov> Co-authored-by: Pete R Jemian <prjemian@gmail.com> Co-authored-by: dependabot[bot] <49699333+dependabot[bot]@users.noreply.github.com> Co-authored-by: Pete R Jemian <prjemian@users.noreply.github.com> Co-authored-by: Mark Wolfman <canismarko@gmail.com>
Fixes a race condition where the SR570 pre-amp gets the sensitivity_value set before the sensitivity_unit, ending up at the wrong gain.
If the gain is set to 1 mA/V in the IOC, and you try to set it to e.g. 200 µA/V, you need to set the unit first then the value since "1" is the only valid "mA/V" value:
✓ 1 mA/V -> 1 µA/V -> 200 µA/V
✗ 1 mA/V -> 200 mA/V (corrects to 1 mA/V) -> 1 µA/V
Solved this problem by using
put_complete=True
for the sensitivity_unit signal.