NI DMM Trigger Functions

Siddhartha Ray edited this page Apr 23, 2021 · 1 revision

Trigger Functions

niDMM_ConfigureTrigger

ViStatus = niDMM_ConfigureTrigger(ViSession Instrument_Handle, ViInt32 Trigger_Source, ViReal64 Trigger_Delay)

Purpose

Configures the DMM Trigger_Source and Trigger_Delay. Refer to Triggering and Using Switches for more information.

Parameters

Input
Name Type Description
Instrument_Handle ViSession Identifies a particular instrument session. You obtain the Instrument_Handle parameter from niDMM_init or niDMM_InitWithOptions. The default is None.
Trigger_Source ViInt32 Specifies the Trigger_Source that initiates the acquisition. The driver sets [NIDMM_ATTR_TRIGGER_SOURCE](javascript:LaunchMergedHelp('dmm.chm', 'dmmcref.chm',
'caNIDMM_ATTR_TRIGGER_SOURCE.html')) to this value. Software configures the DMM to wait until [niDMM_SendSoftwareTrigger](javascript:LaunchMergedHelp('dmm.chm', 'dmmcref.chm',
'cviniDMM_SendSoftwareTrigger.html')) is called before triggering the DMM.

Note  To determine which values are supported by each device, refer to the LabWindows/CVI Trigger Routing section.

Trigger_Delay ViReal64 Specifies the time that the DMM waits after it has received a trigger before taking a measurement. The driver sets the NIDMM_ATTR_TRIGGER_DELAY, which means the DMM waits an appropriate settling time before taking the measurement. On the NI 4060, if you set Trigger_Delay to 0, the DMM does not settle before taking the measurement. The NI 4065 and NI 4070/4071/4072 use the value specified in Trigger_Delay as additional settling time.

Note  When using the NI 4050, Trigger_Delay must be set to NIDMM_VAL_AUTO_DELAY (-1).

Return Value

Name Type Description
Status ViStatus Reports the Status of this operation. To obtain a text description of the status code, call niDMM_GetErrorMessage. To obtain additional information concerning the error condition, use niDMM_GetError.

niDMM_SendSoftwareTrigger

ViStatus = niDMM_SendSoftwareTrigger(ViSession Instrument_Handle)

Purpose

Sends a command to trigger the DMM. Call this function if you have configured either the [NIDMM_ATTR_TRIGGER_SOURCE](javascript:LaunchMergedHelp('dmm.chm', 'dmmcref.chm', 'caNIDMM_ATTR_TRIGGER_SOURCE.html')) or [NIDMM_ATTR_SAMPLE_TRIGGER](javascript:LaunchMergedHelp('dmm.chm', 'dmmcref.chm', 'caNIDMM_ATTR_SAMPLE_TRIGGER.html')) attributes. If the [NIDMM_ATTR_TRIGGER_SOURCE](javascript:LaunchMergedHelp('dmm.chm', 'dmmcref.chm', 'caNIDMM_ATTR_TRIGGER_SOURCE.html')) and/or [NIDMM_ATTR_SAMPLE_TRIGGER](javascript:LaunchMergedHelp('dmm.chm', 'dmmcref.chm', 'caNIDMM_ATTR_SAMPLE_TRIGGER.html')) attributes are set to NIDMM_VAL_EXTERNAL or NIDMM_VAL_TTLn, you can use this function to override the trigger source that you configured and trigger the device. The NI 4050 and NI 4060 are not supported.

Parameters

Input
Name Type Description
Instrument_Handle ViSession Identifies a particular instrument session. You obtain the Instrument_Handle parameter from niDMM_init or niDMM_InitWithOptions. The default is None.

Return Value

Name Type Description
Status ViStatus Reports the Status of this operation. To obtain a text description of the status code, call niDMM_GetErrorMessage. To obtain additional information concerning the error condition, use niDMM_GetError.

niDMM_ConfigureMeasCompleteDest

ViStatus = niDMM_ConfigureMeasCompleteDest(ViSession Instrument_Handle, ViInt32 Meas_Complete_Destination)

Purpose

Specifies the destination of the DMM Measurement Complete (MC) signal. Refer to Triggering for more information.

Parameters

Input
Name Type Description
Instrument_Handle ViSession Identifies a particular instrument session. You obtain the Instrument_Handle parameter from niDMM_init or niDMM_InitWithOptions. The default is None.
Meas_Complete_Destination ViInt32 Specifies the destination of the Measurement Complete signal. This signal is issued when the DMM completes a single measurement. The driver sets the NIDMM_ATTR_MEAS_COMPLETE_DEST attribute to this value. This signal is commonly referred to as Voltmeter Complete.

Note  To determine which values are supported by each device, refer to the LabWindows/CVI Trigger Routing section.

Return Value

Name Type Description
Status ViStatus Reports the Status of this operation. To obtain a text description of the status code, call niDMM_GetErrorMessage. To obtain additional information concerning the error condition, use niDMM_GetError.

niDMM_ConfigureTriggerSlope

ViStatus = niDMM_ConfigureTriggerSlope(ViSession Instrument_Handle, ViInt32 Trigger_Slope)

Purpose

Sets the NIDMM_ATTR_TRIGGER_SLOPE polarity.

Parameters

Input
Name Type Description
Instrument_Handle ViSession Identifies a particular instrument session. You obtain the Instrument_Handle parameter from niDMM_init or niDMM_InitWithOptions. The default is None.
Trigger_Slope ViInt32 Specifies the polarity of the trigger signal on which the measurement is triggered for values of either NIDMM_VAL_POSITIVE or NIDMM_VAL_NEGATIVE. The driver sets the NIDMM_ATTR_TRIGGER_SLOPE attribute to this value.
NIDMM_VAL_POSITIVE 0 The driver triggers on the rising edge of the trigger signal.
NIDMM_VAL_NEGATIVE (default) 1 The driver triggers on the falling edge of the trigger signal.

Return Value

Name Type Description
Status ViStatus Reports the Status of this operation. To obtain a text description of the status code, call niDMM_GetErrorMessage. To obtain additional information concerning the error condition, use niDMM_GetError.

niDMM_ConfigureSampleTriggerSlope

ViStatus = niDMM_ConfigureSampleTriggerSlope(ViSession Instrument_Handle, ViInt32 Sample_Trigger_Slope)

Purpose

Sets the NIDMM_ATTR_SAMPLE_TRIGGER_SLOPE polarity.

Parameters

Input
Name Type Description
Instrument_Handle ViSession Identifies a particular instrument session. You obtain the Instrument_Handle parameter from niDMM_init or niDMM_InitWithOptions. The default is None.
Sample_Trigger_Slope ViInt32 Specifies the polarity of the Trigger signal on which the measurement is triggered for values of either NIDMM_VAL_POSITIVE or NIDMM_VAL_NEGATIVE. The driver sets NIDMM_ATTR_SAMPLE_TRIGGER_SLOPE to this value.
Rising Edge 0 NIDMM_VAL_POSITIVE The driver triggers on the rising edge of the trigger signal.
Falling Edge (default) 1 NIDMM_VAL_NEGATIVE The driver triggers on the falling edge of the trigger signal.

Return Value

Name Type Description
Status ViStatus Reports the Status of this operation. To obtain a text description of the status code, call niDMM_GetErrorMessage. To obtain additional information concerning the error condition, use niDMM_GetError.

niDMM_ConfigureMeasCompleteSlope

ViStatus = niDMM_ConfigureMeasCompleteSlope(ViSession Instrument_Handle, ViInt32 Meas_Complete_Slope)

Purpose

Sets the Measurement Complete signal to either rising edge (positive) or falling edge (negative) polarity.

Parameters

Input
Name Type Description
Instrument_Handle ViSession Identifies a particular instrument session. You obtain the Instrument_Handle parameter from niDMM_init or niDMM_InitWithOptions. The default is None.
Meas_Complete_Slope ViInt32 Specifies the polarity of the signal that is generated. The driver sets NIDMM_ATTR_MEAS_DEST_SLOPE to this value.
Rising Edge 0 NIDMM_VAL_POSITIVE The driver triggers on the rising edge of the trigger signal.
Falling Edge (default) 1 NIDMM_VAL_NEGATIVE The driver triggers on the falling edge of the trigger signal.

Return Value

Name Type Description
Status ViStatus Reports the Status of this operation. To obtain a text description of the status code, call niDMM_GetErrorMessage. To obtain additional information concerning the error condition, use niDMM_GetError.

Table of Contents

Internal Development

Creating and Setting Up a gRPC Server

Server Security Support

Creating a gRPC Client

gRPC Client Examples

Session Utilities API Reference

Driver Documentation

gRPC API Differences From C API

Sharing Driver Sessions Between Clients

C API Docs
NI-DAQmx
NI-DCPOWER
NI-DIGITAL PATTERN DRIVER
NI-DMM
NI-FGEN
NI-FPGA
NI-RFmx Bluetooth
NI-RFmx NR
NI-RFmx WCDMA
NI-RFmx GSM
NI-RFmx CDMA2k
NI-RFmx Instr
NI-RFmx LTE
NI-RFmx SpecAn
NI-RFmx TD-SCDMA
NI-RFmx WLAN
NI-RFSA
NI-RFSG
NI-SCOPE
NI-SWITCH
NI-TCLK
NI-XNET
Clone this wiki locally